JSNM Akita, Japan, 1999 Performance characteristics of PCD system in Axial filter and Open frame
acquisition mode
|
ABSTRACT
This study characterizes the performance of Axial filter and Open frame acquistion mode in PCD system
(Prism-200XP, Picker, USA). Intrinsic spatial resolution, scatter fraction, senistivity, and image quality
is evaluated. Spatial resolution in (FWHM) is 5.7mm at the center for both Axial filter and Open frame.
Scatter fraction is 19% for Axial filter and 27% for the Open frame. Sensitivity measurements of total events
(in cps/ Bq/ ml) for Axial filter and Open frame ar 1.0 and 4.2, respectively. Conclusion: Open frame acquisition
mode could be clinically applicable in low administrative doses or in reducing the scanning time.
|